A low-noise multi-electrode array system for
in vitro extracellular electrophysiology

MED64 System

MED Probe

MED64 Amplifier

MED Connector

Software

High-throughput

Perfusion Accessories

Key Technologies

KEY TECHNOLOGIES

MED64 Key technology
Technical advantages
Patents
History

horizontal line

MED64 Key Technology

Low-impedance planar microelectrodes

The use of MEAs (Multi-electrode arrays) has been growing in popularity over the last several years. Among MEA systems, the key technology unique to the MED64 system is its “low-impedance” planar microelectrode coated with platinum black. This low-impedance planar microelectrode enables recording from diverse preparations including acute slices as well as cultures and explants. It allows beginners to electrophysiology to easily acquire clean, high quality signals and perform sophisticated experiments. The development of the planar microelectrodes for MEAs required solving a major problem. Electrode impedance is inversely proportional to its surface area but surface area cannot be too large. This dilemma was solved by coating the planar microelectrodes with platinum black using special method. This technology increases the surface area 100-200 times (greater than standard plating methods) and reduces the electrode impedance to only 7-10 kohm at 1 kHz (for 50x 50 µm electrode, 40 kohm for 20 x20 µm electrode), which results in following benefits.

  1. Low-noise and stable recording
  2. Superior signal-to-noise ratio
  3. Excellent electrical stimulation
  4. No need for pre-amplifier
Electrode 2
electrode surface 3d
 
Close up views of electrode's surface

 
Comparison Electrode
Impedance of Pt. black electrode on the MED probe (Red: 50 x 50 µm, Green: 20 x20 µm) compared with completely flat ITO electrode (Blue: 50 x50 µm)
 

Back to TOP

Technical Advantages

1. Low noise

There are two types of noise to be considered in a typical electrophysiology experiment, exogenous noise and Johnson noise. Both types of noise decrease as the electrode-impedance becomes smaller.

Low exogenous noise

One representative example of exogenous noise is hum noise. The lower the electrode impedance, the less susceptible the system is to exogenous noise. Extracellular recordings with conventional glass electrodes often require a special environment such as Faraday cage and a vibration isolation table. The MED64 system does not usually require a Faraday cage but can be installed on a stable table. Experiments are conducted everyday under stable low-noise level.

experiment
Fig.1: Experiment using the MED64 system
           MED probe and connector are placed in the incubator.

Low Johnson noise

Johnson noise is intrinsic to the system itself and can be determined by the amplitude of the baseline noise. Unlike exogenous noise, the level of Johnson noise correlates with the electrode’s impedance and can NOT be improved by users. Low electrode-impedance will decrease the Johnson noise. The measured noise level (Vn) of the MED64 system (50 x 50 µm electrodes) is as low as 2-3µV RMS at 0.1-10kHz. (Noise levels below 1µV RMS can be achieved applying the 3kHz low-pass 9 pole Bessel filter included in the MED64 Mobius software).

 


(0.01 mV, 10 msec /div.)



(0.02mV, 20 msec / div.) 


Fig.2: noise level obtained with saline (Bandwidth of 0.1-10kHz)

2. High signal to noise ratio

The signal to noise ratio is the ratio of signal compared to the noise. Obviously, higher quality of signals could be recorded with higher signal to noise ratio. High signal to noise ratio is achieved by enlarging signal level or lowering noise. With planner microelectrodes at Multi-electrode array, it is impossible to enlarge the signal level. (In order to enlarge signal level, the electrode which point’s size is around several micro meters needs to be inserted deep into the middle of slices.) Thus, lowering the noise level is the only method to achieve high signal to noise ratio. The MED64’s Johnson noise is very small due to its low-impedance microelectrodes High-quality signals can be recorded even from acute slices.

Click here for Publication list of acute slices


Fig.5: Raw data from acute hippocampus slice (0.5mV, 10ms/div)

3. Effective stimulation and recording of good electrically evoked responses

One of the capabilities of the MED64 system excels at is recording electrically-evoked activity. As mentioned previously, the Pt. black electrode has a large surface area which is associated to high capacitance and low impedance. As a result, high current stimulation can be applied and good evoked responses are recorded without interference of stimulus artifacts.

High current stimulation

Fig.6 shows the mechanism of extracellular electrical stimulation by planar microelectrodes in the MED64 system. The stimulus current flows from the selected electrode to reference electrodes. This current changes the field potentials, which results in the hyperpolarization or depolarization of cellular membranes. The voltage changes in the field potentials (Vf) corresponds to the stimulator output current (Is) as seen in the right figure of Fig.6.

Stimulation Mechanism


Ce: Capacitance of electrical double-layer capacitor (electrode)
Re: Resistance of electrode
Rs: Resistance of solution with slice
Is: Stimulator output current
Vs: Stimulator output voltage
Vc: Voltage of Ce
Vf: Voltage of Rs (Voltage change in the field)

 

Fig.6. Stimulation by the MED64's planar microelectrode

The applied stimulus current charges the “electrical double-layer capacitor” (Ce) which is formed at the interface of electrode, which causes the voltage of Ce (Vc) to change. However, when the absolute voltage (Vc) rises above a certain limit (around 1V for materials which are usually used for MEAs), the electrode will reach electrolysis and release H2 bubbles. This causes the stimulation efficacy to drop dramatically, introduces extremely large stimulus artifacts, harm cells, and damages electrodes. Thus, the stimulus current (Is) should not go beyond this level.

The (Vc) is defined by

Vc =
Is x t   Vc: Voltage of Ce / Is: Stimulation output current
  Ce   T: Pulse duration / Ce: Capacitance of electrical double layer capacitor (electrode)

The higher the electrode’s capacitance (Ce) is, the lower its voltage (Vc) is. This also means that the higher the capacitance of the electrode is, the greater the tolerance to current (Is) is.

The capacitance of MED64 probe’s microelectrodes coated with Pt. black is as high as 50,000 pF for 50 x 50µm electrodes (22,000 pF for 20 x 20µm electrode). The value is much higher than any other material used for MEAs. For example, the (Vc) at the stimulus current of 20µA for the MED64’s microelectrode is much lower than the one for another material. (See Fig.7). This high capacitance electrodes enables stimulation with current amplitudes as high as 200µA/0.2ms.

 
Fig.7: Comparison of (Vc) for different types of
microelectrodes when 20 mA stimulus current is applied

Small stimulus artifact

When the planar electrodes at MEAs is used for electrical stimulation, the pulses of stimulus current are recorded at other channels due to crosstalk. This results in the so-called “stimulus artifact”. The stimulus artifact can persist even after stimulus pulse returns to zero volts and interfere with the recording of signals as seen in figure 8.

This problem is mainly due to following factors:

  1. Usually, the maximum input voltage for the amplifier is set to several milli-volts. However, when recording with MEAs, the amplitude of the incoming stimulus artifact exceeds these levels, which causes the amplifier to saturate momentarily making data acquisition impossible. This delay due to the amplifier’s momentary saturation becomes larger in proportion to the stimulator output voltage (Vs).
  2. The electrode often needs longer time to discharge than to charge. (Refer to the (Vc) at Fig.9) This discrepancy becomes greater as the voltage of Ce (Vc) becomes larger.
 
Artifact
Fig.8: Prolonged stimulus artifacts

As described in the previous chapter, the (Vc) at the MED64 electrodes can be kept very small because of its high capacitance. Moreover, the (Vs), can be kept small by lowering the electrode’s impedance (Refer to the (Vs) and (Vc) at Fig.9) As a result, the prolongation of the stimulus artifact can be kept small thanks to the low-impedance and high-capacitance electrodes at the MED64 system.

Fig.9: Comparison of voltages between two different electrodes. (Vs) is smaller for lower-impedance electrode and (Ce) is smaller for higher capacitance electrode

 

The stimulus artifact can return to 0 volt within 0.5 m seconds at non-stimulated channels with the MED64 system even with current stimuli as high as 100µA as seen in figure 10, and rarely interferes with recorded signals.


(0.1mV, 20ms /div.)
 
(0.1mV, 0.2ms  /div)

Fig.10: The stimulus artifact with the MED64 system when a 100 µA current pulse is applied.

Hardware-implemented approaches such as electronic blanking circuitry can solve the problem only due to the first factor by amplifier’s saturation. Low-impedance and high-capacitance electrodes can be a good approach to minimize the duration of the stimulus artifact.

Finally, the underlying reason for this artifact problem is capacitive crosstalk. Thus, it is the important and fundamental solution to minimize the cross talk. Lowering the electrode impedance is one way of achieving this.

4. No need for pre-amplifier.

Another benefit which is unique to the MED64 system is that the preparation can be placed in a humidified incubator during recordings, which works very well for long-term recording and precise control of the temperature. This feature is also enabled by its low-impedance electrodes. With higher impedance electrodes, a pre-amplifier needs to be located close to the electrode to minimize exogenous noise and prevent signal attenuation, under some circumstances. However, it is not necessary with the MED64’s low-impedance microelectrode. The MED probe can be connected to the amplifier via a two meter long cable and the MED connector with no active circuitry. This system configuration allows you more flexibility for your experiments.

Back to TOP


MED64 Patents

The novelty and originality of the research and development carried out at the Corporate Research Division of Panasonic. in Osaka, Japan, has been recognized with the award of the following  patents in the US, Europe, Japan, Korea and Taiwan. The technology described therein has been incorporated into the components of the MED64 system, making these the only legitimate products with the large area microelectrodes indispensable for effective stimulation.

MED System

USA
USP 5,563,067, October 1996
Europe
EP00689051B1, November 2002
Japan
JP3204875, June 2001
Korea
KR0150390B1, October 1998

 

MED Probe

USA
USP 6,151,519, November 2000
USP 6,132,683, October 2000
USP 5,810,725, September 1998
Europe
EP01040345B1, March 2006
Japan
JP2949845, July 1999
JP3101122, August 2000
JP3193471, May 2001
Taiwan
TW128335, July 2001

Back to TOP


History

Among various studies pursuing multi-site recording from brain circuits, planar microelectrode arrays have been shown to be appropriate for recording responses of dissociated cultured neurons obtained from fetal and neonatal rat brains (Pine, 1980; Gross et al., 1982)*. However, it was recognized that the networks of dissociated cultured neurons might respond differently from the neural organization of intact brains.  

Gross, G. W., Williams, A. N. and Lucas, J. M., J. Neurosci. Meth., 1982; 5: 13-22.

 

Application of planar microelectrode arrays to acute brain tissue was reported by Novak and Wheeler in 1988, where they showed that their array could both stimulate and record from multiple sites from hippocampal slice preparations. Although their work definitely showed that planar microelectrode arrays could be useful for studies of brain tissue, stimulation by their array electrodes seemed somewhat ineffective, and their recordings were limited to 2-4 hours.

 

Novak, J. L. and Wheeler, B. C., J. Neurosci. Meth., 1988; 23: 149-159

The MED systems approach to achieving both effective stimulation and long periods of recording time is based on two key properties:
1. planar electrodes with a large area; and
2. the coating on the surface of the array.

 

An array of 64 planar microelectrodes was developed, each having a size of 50 x 50 µm, arranged in an 8 by 8 pattern (the MED probe, USP 5,810,725). The material of the insulating layer of the MED probe, which directly contacts the brain tissue, was carefully chosen. The coating material and method of preparation for the surface of the MED probe were also studied. These technologies, together with electronics designed exclusively for the MED probe (USP 5,563,067), enabled multi-channel recordings of field EPSPs induced by stimulation of array electrodes for more than six hours (Oka et al.). Oka, H., Shimono, K., Ogawa, R., Sugihara, H., & Taketani, M., J. Neurosci. Meth., 1999 ; 93: 61-67

Back to TOP

 

 

 
 

HOME    PRODUCTS    APPLICATIONS     PUBLICATIONS    SUPPORT   CONTACT  ABOUT US
Copyright: 2010 Alpha MED Scientific Inc
.
Alpha MED Scientific Inc., 209 Saito-incubator, 7-7-15, Saito-asagi, Ibaraki Osaka, 567-0085, Japan