A low-noise multi-electrode array system for
in vitro extracellular electrophysiology

MED64 System

MED Probe

MED64 Amplifier

MED Connector

Software

High-throughput

Perfusion Accessories

Key Technologies

MED PROBE

MED Probe

 

Multi-electrode array with the industry’s lowest impedance electrodes.

Acute or cultured biological preparations are placed or grown directly on a grid of 64 planar microelectrodes, with the capability for stimulation and signal recording.

The standard MED probe has 64 planar microelectrodes arranged in an 8 x 8 grid embedded in the center of a transparent glass plate. The surrounding glass or plastic cylinder makes the MED probe a self-contained recording chamber. Four models with inter-electrode spacings of 100, 150, 300, and 450µm enable detailed evaluation of network interactions across a sample. Each electrode is 50µm x 50µm in size for 150, 300, and 450µm spacing configuration, or 20µm x 20µm for 100µm spacing configuration.

Also available are probes featuring an electrode pattern specific to hippocampus anatomy and a hexagonal pattern, as well as 2- and 4-well MED probes for increasing throughput.

The micro-planar electrodes patterned on the MED probe have the industry’s lowest impedance (10kΩ at 1 kHz for 50μm electrode) which results in superior signal-to-noise ratios.
(Learn MORE about the technology and related benefits)

 

Standard 8x8 Arrays

   
Interpolar Distance of Electrodes
 

Product Number

Array size
Electrode size
Recording Electrodes
Reference Electrodes
Chamber Depth
MED-P2105
0.7x0.7mm
20x20 µm
100 µm
8.5 mm
5 mm
MED-P210A
0.7x0.7mm
20x20 µm
100 µm
8.5 mm
10 mm
MED-P5155
1 x 1 mm
50x50 µm
150 µm
8.5 mm
5 mm
MED-P515A
1 x 1 mm
50x50 µm
150 µm
8.5 mm
10 mm
MED-P5305
2 x 2 mm
50x50 µm
300 µm
9.2 mm
5 mm
MED-P530A
2 x 2 mm
50x50 µm
300 µm
9.2 mm
10 mm
MED-P5455
3 x 3 mm
50x50 µm
450 µm
10.2 mm
5 mm
MED-P545A
3 x 3 mm
50x50 µm
450 µm
10.2 mm
10 mm

 

MED-P515A Close Up View
MED-P515A Close Up View

MED-P210# (0.7mm)

Mouse Hippocampus CA1

MED-P515# (1mm)

Mouse Hippocampus

MED-P515# (1mm)

Rat Hippocampus CA1

MED-P530# (2mm)

Rat Hippocampus

MED-P545# (3mm)

Rat Hippocampus and Peripheral area

Hippocampal Pattern Arrays (CA1, CA3, & DG)

Interpolar Distance of Electrodes
Parts number
Electrode size
Recording electrodes
Reference electrodes
Chamber Depth
MED-P50015
For Rat
ø 50 µm
150 µm
8.5 mm
5 mm
MED-P5001A
For Rat
ø 50 µm
150 µm
8.5 mm
10 mm
MED-P50025
For Mouse
ø 50 µm
150 µm
8.5 mm
5 mm
MED-P5002A
For Mouse
ø 50 µm
150 µm
8.5 mm
10 mm

MED-P5001A
Micrograph of rat hippocampal slice and arrangement of microelectrodes. Click to zoom.

MED-P5002A
Micrograph of rat hippocampal slice and arrangement of microelectrodes. Click to zoom.

Hexagonal Arrays (61 Electrodes)

Interpolar Distance of Electrodes
Product Number
Electrode size
Recording electrodes
Reference electrodes
Chamber Depth
MED-P2H075
ø 50 µm
70 µm
8.5 mm
5 mm
MED-P2H07A
ø 50 µm
70 µm
8.5 mm
10 mm

MED-P2H07A
Close up view and arrangement of microelectrodes.
Click to zoom.

MED probes for Parallel Sample Processing

2-Sample MED probe

2- Sample MED Probe (2-well type, MED-P5D15A) (left) and schematic diagram of microelectrode pattern (right).

Product Number
Number of wells
Electrode size
Inter-polar distance
(The smallest)
Distance between
electrode blocks
Chamber
Depth
Partition-
height
MED-P50035
1
ø 50 µm
150 µm
12 mm
5 mm
NA
MED-P5003A
1
ø 50 µm
150 µm
12 mm
10 mm
NA
MED-P5D15A
2
ø 50 µm
150 µm
NA
10 mm
10 mm
MED-P5D15B
2
ø 50 µm
150 µm
NA
10 mm
5 mm

4-Sample MED probe

4- Sample MED Probe (4-well type, MED-P5FF15) (left) and schematic diagram of microelectrode pattern (right)

Product Number
Number of
wells
Electrode
Size
Inter-polar distance
(The smallest)
Distance between
electrode blocks
Chamber
Depth
Partition-
height
MED-P50045
1
ø 50 µm
150 µm
5.5
5 mm
NA
MED-P5004A
1
ø 50 µm
150 µm
5.5
10 mm
NA
MED-P5DF15
2
ø 50 µm
150 µm
NA
10 mm
10 mm
MED-P5FF15
4
ø 50 µm
150 µm
NA
10 mm
10 mm

Specifications

Substrate
Glass substrate: 50 x 50 x 0.7mm
Glass cylindrical chamber: ID=22, OD=25mm
Conducting layer: Indium tin oxide (ITO) ( 0.15µm)
Insulation layer: Polyacrylamide (1.5µm)
Recording / Stimulating electrodes
Material ITO + Platinum Black
Impedance < 22kΩ (Typical:10Ω) for 50µm electrodes
< 40kΩ (Typical : 30Ω) for 20µm electrodes
(1kHz, 50mV applied sinusoidal wave)
Reference electrodes
Material ITO + Platinum Black
Size 200 x 200µm
ø 100µm x 4 (MED-P2H07#, MED-P500##, MED-P5D###, MED-P5F###)

 

 
 

HOME    PRODUCTS    APPLICATIONS     PUBLICATIONS    SUPPORT   CONTACT  ABOUT US
Copyright: 2010 Alpha MED Scientific Inc
.
Alpha MED Scientific Inc., 209 Saito-incubator, 7-7-15, Saito-asagi, Ibaraki Osaka, 567-0085, Japan