A low-noise multi-electrode array system for
in vitro extracellular electrophysiology

MED64 System

MED Probe

MED64 Amplifier

MED Connector

Software

High-throughput

Perfusion Accessories

Key Technologies

MED64 AMPLIFIER

MED64 Head amplifier (top) and Main amplifier (bottom)

  • Bandwidth of 0.1Hz -10 kHz allows users to record several types of extracellular potentials.
  • Low-cut filter (High pass filter) with 0.1, 1, 10, and 100Hz cutoff.
  • High cut filter (Low pass filter) with 1, 2.5, 5, 7.5, and 10kHz cutoff.
  • Built-in 2-channel stimulator allows users to program and deliver complex stimulation sequences
  • independently through any 2 electrodes.
  • USB connection to computer.

 

MED64 Head Amplifier
(MED-A64HE1)

Amplifier

Number of channels

64

Gain

x 10

Bandwidth

0.1Hz-100 kHz (+0dB to -3dB)

Input impedance

100MΩ

Output impedance

 

10kΩ

Stimulator

Number of channels

2

Output format:

Constant current

Maximum input voltage

+/-4V

Maximum output current

 

+/-200µA

General

Power supply

DC +/- 12V

Weight

6.6Kg

Dimensions

 

W430 x H74 x D437mm

Power supply unit

Input

AC 100-240V (50-60Hz)

Output

DC +/- 12V

 

MED64 Main Amplifier
(MED-A64MD1)

Amplifier
Number of channels 64
Gain 1-217
Bandwidth 0.1Hz-10 kHz (+0dB to -3dB)
Analog high cut filter 1, 2.5, 5, 7.5, 10kHz (-6dB/oct)
Analog low cut filter 0.1, 1, 10, 100Hz (-12dB/oct)

Input impedance

 

100 MΩ
Digitizer
Resolution 16 bit
Sampling rate 20kHz

Output:

 

USB
General
Power supply DC +/-12V
Weight 5.9Kg

Dimensions

 

W430 x H74 x D437mm
Power supply unit
Input AC 100-240 V (50-60Hz)
Output DC +/- 12V
 
 
 